ANSI/AAMI ST108:2023 Tables 1–3 · Live Lab CoA Grader
ST108 Water Result Quick-Check
Paste readings from your hospital lab Certificate of Analysis to grade Critical Water, Utility Water, and Steam Condensate against ANSI/AAMI ST108:2023 action limits—and hand off flagged excursions directly into your Annex G plan.
Critical Water (RO/DI Final Rinse)Table 2
Extensively treated RO/DI water for final rinse of critical & semi-critical medical devices.
PASS
µS/cm
Limit: ≤ 10 µS/cmAlert > 5 µS/cm
PASS
EU/mL
Limit: ≤ 0.25 EU/mLAlert > 0.06 EU/mL
ST108 Table 2 allows <10 EU/mL for general critical water; WaterShield flags >0.25 EU/mL clinical action limit.
PASS
CFU/mL
Limit: ≤ 10 CFU/mLAlert > 1 CFU/mL
PASS
pH units
Limit: 5–7.5 pH units
PASS
mg/L
Limit: ≤ 1 mg/LAlert > 0.5 mg/L
PASS
mg/L
Limit: ≤ 0.1 mg/LAlert > 0.05 mg/L
Live ST108 Compliance Verdict
In-Spec (ANSI/AAMI ST108 Pass)
All entered critical water analytes satisfy ANSI/AAMI ST108:2023 thresholds with no SPC alert trends. Import this profile into the ST108 Wizard to automate monthly lab CoA ingestion and Joint Commission binder exports.